Information

  • Audit Title

  • Document No.

  • Client / Site

  • Conducted on

  • Prepared by

  • Location
  • Personnel

Interferences

  • Does the data system correct for Isobaric elemental interferences?

  • When memory interferences are noted, is the system flushed with a rinse blank to minimize build-up?

Sample Collection, etc.

  • Has the sample pH been checked; to ensure proper preservation techniques?

  • Sample acidified to <pH2 with nitric acid?

Reagents and Standards

  • Are Ultra High purity reagents used for analysis?

  • D.1.4.b.2 Is the quality of water sources monitored and documented to meet method specified requirements? (ASTM Type I ?)

  • Are fresh multi-element calibration standards prepared every two weeks?

  • Internal Standard

  • Is the instrument calibrated by using an internal standard technique?

  • What are the Internal standards? (scandium, yttrium, indium, terbium and bismuth)

  • Are a minimum of three internal standards used (10.3)

QC

  • 5.6.2.b, C.1 Does the laboratory management maintain records to assure that all technical laboratory staff and/or work cells have demonstrated and documented initial and ongoing proficiency in the activities for which they are responsible?

  • Has the upper limit of the linear dynamic range been established?

  • Blanks

  • Calibration blank - Consists of 1% (v/v) nitric acid in reagent grade water.

  • Laboratory reagent blank (LRB) - Must contain all the reagents in the same volumes as used in processing the samples. The LRB must be carried through the same entire preparation scheme as the samples including digestion, when applicable.

  • D.1.1.a Is the method blank processed along with and under the same conditions as the associated samples include all steps of the analytical procedure?

  • Rinse blank - Consists of 2% (v/v) nitric acid in reagent grade water. Rinse blanks should be used to flush the system between blanks, samples, standards and QC samples.

  • Tuning

  • D.1.5.c Does the laboratory document acceptance criteria for mass spectral tuning?

  • Tuning solution

  • Be, Mg, Co, In, and Pb; 100 ug/L each in 1% v/v HNO3

  • Prior to analysis each day, does the lab conduct mass calibration and resolution checks using the tuning solution. Resolution at low mass is indicated by magnesium isotopes 24, 25, and 26. Resolution at high mass is indicated by lead isotopes 206, 207, and 208.

  • Is the mass calibration adjusted if it has shifted by more than 0.1 amu from unit mass.

  • Is instrument stability demonstrated by running the tuning solution a minimum of five times with resulting relative standard deviations of absolute signals for all analytes of less than 5%.

  • Laboratory Fortified Blank /LCS & QCS

  • Is a QCS analyzed initially and periodically of calibration standards or stock standard solutions in order to verify instrument performance? Is a fresh solution created quarterly or more frequently as needed?

  • D.1.1.b.1 Is a LCS performed at a frequency of 1 per preparation batch of per matrix type, except for analytes for which spiking solutions are not available?

  • The LFB must be carried through the same entire preparation scheme as the samples including sample digestion, when applicable.

  • Are upper and lower control limits for the LFB 85-115%?

  • Is a calibration blank analyzed at the beginning of the run and after each 10 samples, and at the end of the run?

  • Is a calibration check standard analyzed at the beginning of the run and after each 10 samples, and at the end of the run?

  • The analysis of all analytes within the calibration check standard solutions must be within ±10% of calibration.

  • If the calibration cannot be verified within the specified limits, the instrument must be recalibrated.

  • D.1.1.c Does the laboratory document procedures for determining the effect of the sample matrix on method performance? (Matrix Spike) Note: The spiking frequency is 10% of samples.

  • Are the acceptable recovery ranges for the LFM/MS 70-130%?

  • Is the acceptable response for internal standards within 60-125% of the original response of the calibration blank?

Calibration and Standardization

  • At minimum is a blank and a standard used for calibration?

  • 5.9.4.2.1.f If the results of samples are not bracketed by the initial calibration, are the results reported as having less certainty (defined qualifiers, flags, or explanation in the case narrative)?

  • Are 3 integrations averaged for calibration and data reporting?

  • 5.9.4.2.1.d Are all initial calibrations verified with a standard obtained from a second source or lot that is traceable to a national standaed when available? (if the lot can be demonstrated from the manufactureras prepared independently from other lots)

  • 5.9.4.2.1.e Is the criteria for the acceptance of an initial calibration established (QCS standard ± 10% Section 9.2.3)

  • 5.9.4.2.1.g Are corrective actions performed if the results of the initial calibration are outside of established acceptance criteria?

  • 5.9.4.2.1.g Is data associated with unacceptable initial instrument calibration not reported?

  • 5.9.4.2, 5.9.4.2.2 Is a continuing instrument calibration verification used to confirm the continued validity of the initial calibration with each sample batch?

  • Continuing calibration is done after every 10 samples (9.3.4) The results of the CCV are ± 10% for each analyte of interest.

  • If the results are between ±10-15% the previous 10 samples can be used, but the instrument must be recalibrated.

  • If the results are >± 15% the previous 10 samples must be re-analyzed.

Procedure

  • Aqueous Sample Prep – Dissolved Analytes (ground and surface waters)

  • Sample volume? (20 mL)

  • Is 0.4 mL of 1:1 HNO3 added to 20 mL of filtered, acid preserved sample? (or equivalent ratio to produce 1% v/v acid/sample) Matrix Matching.

  • Aqueous Sample Prep – Total Recoverable Analytes (drinking water)

  • For unfiltered samples < 1 NTU, is the above procedure used?

  • Sample volume? (100 mL)

  • Digestion:

  • -          Add 2mL 1:1 HNO3 and 1mL 1:1 HCl.

  • -          Heat on hot plate at 85C.

  • -          Reduce volume to 20mL.

  • -          Takes 2 hours for reduction. Do not boil sample.

  • -          Cover beaker with watch glass and reflux sample for 30 minutes.

  • -          Cool sample.

  • -          Transfer to 50mL flask, dilute to volume with reagent water.

  • -          Let sit overnight.

  • -          Centrifuge if solids show.

  • -          Adjust chloride concentration.

  • Solid Sample Prep – Total Recoverable Analytes

  • Sample weight for percent solids/moisture determination?

  • < 35 % moisture use ~ 20 g

  • > 35 % moisture use ~ 50 – 100 g

  • Is a percent solids calculation used for reporting results on a dry weight basis?

  • Digestion:

  • -          Weigh 1.0 +/- 0.01 g of dried sample

  • -          Add 4 mL 1:1 HNO3 and 10 mL 1:4 HCl.

  • -          Cover with watch glass

  • -          Heat on hot plate at 85C (covered sample refluxes at 95C).

  • -          Reflux for 30 min. Avoid vigorous boiling.

  • -          Cool sample.

  • -          Transfer to 100 mL flask, dilute to volume with reagent water.

  • -          Let sit overnight.

  • -          Centrifuge if solids show.

  • -          Adjust chloride concentration.

NELAC Related

  • D.1.4.a, 5.9.2.a-c Is the source of standards traceable to national standards

  • D.1.6.a Does the laboratory assure that the test instruments consistently operate within the specifications required of the application for which the equipment is used?

  • D.1 Are all quality control measures assessed and evaluated on an on-going basis?

  • D.1 Are the quality control protocols specified by the laboratory’s method manual followed by all analysts?

  • 5.9.4.2.1.b, 5.9.4.2.2.c Are sufficient raw data records retained to permit reconstruction of the initial and continuing calibrations using as appropriate, but not limited to:

  • a.___ Calibration date

  • b.___ Test method

  • c.___ Instrument

  • d.___ Analysis date

  • e.___ Each analyte name

  • f.___ Analyst's initials or signature

  • g.___ Concentration and response

  • h.___ Response

  • i.___ Calibration curve or response factor, or

  • j.___ Unique equation or coefficient used to reduce instrument responses to concentration.

  • 5.9.4.1.d Prior to use on each working day, are balances, ovens, refrigerators, freezers, incubators and water baths checked with NIST traceable references (where available) in the expected use range?

  • 5.9.2.a Are measurements made by the labs traceable to national standards of measurement where available?

  • 5.10.5.a Does the laboratory retain records for all standards, reagents and media including manufacturer/vendor, the manufacturer’s Certificate of Analysis or purity (if supplied), date of receipt, recommended storage conditions, and an expiration date after which the material shall not be used unless its reliability is verified by the laboratory?

  • 5.9.4.1.c Are all raw data records retained to document equipment performance? (i.e. maintenance logs)

  • 5.9.4.1.e Are mechanical volumetric devices, including burettes, checked for accuracy on at least a quarterly use basis?

  • 2.5 Do the laboratory's management and all analysts ensure that all PT samples are handled (i.e., managed, analyzed, and reported) in the same manner as real environmental samples utilizing the same staff, methods as used for routine analysis of that analyte, procedures, equipment, facilities, and frequency of analysis?

  • Assessor

  • Lab Representative

The templates available in our Public Library have been created by our customers and employees to help get you started using SafetyCulture's solutions. The templates are intended to be used as hypothetical examples only and should not be used as a substitute for professional advice. You should seek your own professional advice to determine if the use of a template is permissible in your workplace or jurisdiction. You should independently determine whether the template is suitable for your circumstances.