AED Checklist Procedure
-
Perform the following procedure each month (28 days):
-
Open the AED lid.
-
Wait for the AED to indicate status. Observe the STATUS INDICATOR change from GREEN to RED during its power up self-test. After approximately 5 seconds, verify that the STATUS INDICATOR returns to GREEN.
-
Check the expiration date on the pads.
-
Check the gauge on the front panel to ensure the battery has sufficient power. Replace the battery if the indicator is RED.
-
Listen for the voice prompts. Additionally, check that the display shows text prompts that correspond to the audio.
-
Close the lid and verify that the STATUS INDICATOR changes from GREEN to RED during its power down self-test. After approximately 5 seconds, verify that the STATUS INDICATOR returns to GREEN.
-