AED Checklist Procedure

  • Perform the following procedure each month (28 days):

  • Open the AED lid.

  • Wait for the AED to indicate status. Observe the STATUS INDICATOR change from GREEN to RED during its power up self-test. After approximately 5 seconds, verify that the STATUS INDICATOR returns to GREEN.

  • Check the expiration date on the pads.

  • Check the gauge on the front panel to ensure the battery has sufficient power. Replace the battery if the indicator is RED.

  • Listen for the voice prompts. Additionally, check that the display shows text prompts that correspond to the audio.

  • Close the lid and verify that the STATUS INDICATOR changes from GREEN to RED during its power down self-test. After approximately 5 seconds, verify that the STATUS INDICATOR returns to GREEN.

  • Signature

The templates available in our Public Library have been created by our customers and employees to help get you started using SafetyCulture's solutions. The templates are intended to be used as hypothetical examples only and should not be used as a substitute for professional advice. You should seek your own professional advice to determine if the use of a template is permissible in your workplace or jurisdiction. You should independently determine whether the template is suitable for your circumstances.