AED Checklist Procedure

Perform the following procedure each month (28 days):

Open the AED lid.

Wait for the AED to indicate status. Observe the STATUS INDICATOR change from GREEN to RED during its power up self-test. After approximately 5 seconds, verify that the STATUS INDICATOR returns to GREEN.

Check the expiration date on the pads.

Check the gauge on the front panel to ensure the battery has sufficient power. Replace the battery if the indicator is RED.

Listen for the voice prompts. Additionally, check that the display shows text prompts that correspond to the audio.

Close the lid and verify that the STATUS INDICATOR changes from GREEN to RED during its power down self-test. After approximately 5 seconds, verify that the STATUS INDICATOR returns to GREEN.

Signature
Please note that this checklist is a hypothetical example and provides basic information only. It is not intended to take the place of, among other things, workplace, health and safety advice; medical advice, diagnosis, or treatment; or other applicable laws. You should also seek your own professional advice to determine if the use of such checklist is permissible in your workplace or jurisdiction.