INSPECTION REPORT FOR BODY-PROTECTED ELECTRICAL AREAS

CVSG ASSET SERVICES PTY LTD
ABN 62 128 821 698

Conducted in accordance with:
AS/NZS 3003:2011 ELECTRICAL INSTALLATIONS - PATIENT AREAS

INSPECTION DETAILS
Select date

NAME OF FACILITY?

LOCATION OF TEST - ROOM NO.?

ASSET BARCODE AVAILABLE?

INSPECTION

All electrical equipment & fittings in good condition?

LABELLING

Circuit identification visible and intact on each?

Protected points labelled as to which RCD protects them?

Circuit identification visible and intact on each LIM?

Protected points labelled as to which LIM monitors them?

Overload monitors labelled with alarm setting?

Socket-outlets protected by RCD labelled ‘RCD PROTECTED’?

Socket-outlets protected by Isolating transformers labelled ‘ISOLATING TRANSFORMER PROTECTED’?

Socket-outlets intended for cleaning purposes labelled ‘CLEANING PURPOSES ONLY’?

Body-protected electrical area sign:
Still visible.
Labelled with current inspection date.
Labelled with identity of authorized person.
Labelled with next test date.

SOCKET OUTLETS

Socket-outlets coloured according to type of supply

If RCD protected, power-available indicator operational?

Number of socket-outlets and permanently wired electrical equipment protected by RCD does not exceed 12?

On/off luminous indicators operational, if applicable?

Socket-outlets double-poled if protected by a transformer-isolated supply?

All socket-outlets below 2.3 m protected by LPD?

Confirm each socket-outlet trips the RCD when fault applied?

Confirm each socket-outlet causes a response on LIM when a fault is applied?

EARTHING

Confirm that the earth pin of all socket-outlets below 2.3 m are earthed to less that 1 Ω?

Confirm that all socket-outlets used by electrical appliances accessible below 2.3 m are earthed to less that 1 Ω?

Confirm that all permanently wired appliances with conductive parts below 2.3 m are earthed to less than 1 Ω?

RCD SENSITIVITY (AUS)

Confirm that the RCD trips at a value greater than 4.5 mA and less than 10.5mA.

Record the sensitivity for each RCD (mA)

RCD TRIPPING TIMES (AUS)

Confirm this by repeating the test at least 2 times.

Record tripping time for each RCD (ms)

RCD SUPPLY AVAILABLE INDICATOR LIGHTS

If applicable, confirm that they are operational.

AUDIO VISUAL RCD TRIP INDICATORS

If applicable, confirm that they are operational.

PROSPECTIVE HAZARD CURRENT (APPLICABLE ONLY TO LIMS)

Confirm that the hazard current is below 2 mA.

Record hazard current (mA)

OVERLOAD MONITORS (APPLICABLE ONLY TO LIMS)

Confirm that a load of 110–120% of transformer rated current causes the alarm to activate in 3–5 s

UNINTERRUPTIBLE POWER SUPPLIES (UPS)

Confirm that UPS status indicator performance is routinely verified.

INSPECTION COMPLETE

INSTALLATION COMPLIANT WITH AS/NZS 3003:2011?

Add signature
Please note that this checklist is a hypothetical example and provides basic information only. It is not intended to take the place of, among other things, workplace, health and safety advice; medical advice, diagnosis, or treatment; or other applicable laws. You should also seek your own professional advice to determine if the use of such checklist is permissible in your workplace or jurisdiction.