LIOM inspection

  • Fittings and fixtures in good condition

  • Is the overload monitor seperate from the Line Isolation Monitor?

  • Labelling in accordance with AS/NZS 3003<br>(A) associated circuit breaker and its distribution panel<br>(B) unique Identifier e.g. "LIM X"<br>(C) Over current Alarm Set Point

  • Push button PHC test

  • Enter PHC results here

Isolation Transformer inspection

  • Fittings and fixtures in good condition?

  • Isolation labelling<br>(A) Circuit breaker and Distribution panel<br>(B) Unique identifier e.g. "LIM X"<br>(C) Manufacturers label intact

  • Sufficient ventilation

  • Isolation transformer is clean and free from dust build up?

  • Circuit breaker rating and type correct for transformer size?<br>Refer to manufacturers specifications

Load Bank test

  • LIOM overload alarm activates when rated load is applied

  • LIOM overload alarm activates with in 3-5 seconds of the load being applied

  • Enter load bank results here LIM number / Overload current / Alarm time

  • Is the installation compliant with AS/NZS 3003:2011 & AS/NZS 4510:1990?

  • Add signature

The templates available in our Public Library have been created by our customers and employees to help get you started using SafetyCulture's solutions. The templates are intended to be used as hypothetical examples only and should not be used as a substitute for professional advice. You should seek your own professional advice to determine if the use of a template is permissible in your workplace or jurisdiction. You should independently determine whether the template is suitable for your circumstances.