Title Page
-
Site conducted
-
• Measuring distance from the sources: 1m, i.e. appr. 60 cm from the surface of the radiation unit.
-
• The measurement is performed in accordance with NCRP 102
-
• Use drawing 43824 for the LGK 23016, 43834 for LGK 23004 and LGK model C and 1002511 for LGK Perfexion as instruction for where to measure.
-
• The result to be in accordance with NCRP 102 (average below 20 μSv/h, maximum 100 μSv/h).
-
• If the measurement is done on LGK 23004, the plastic hood must be removed.
-
• A calibrated instrument shall always be used for the measurement.
-
Conducted on
-
Prepared by
-
Location
Site and Instrument data.
-
LGK Serial number
-
Total Activity (TBq)
-
Reference Date
-
Picture of activity sign on LGK
-
Instrument type
-
Instrument serial number
-
Instrument calibration date
-
Picture of calibration label
Measurements
-
Point 1
-
Point 15
-
Point 3
-
Point 19
-
Point 11
-
Point 23
-
Point 7
-
Point 16
-
Point 4
-
Point 20
-
Point 12
-
Point 24
-
Point 8
-
Point 17
-
Point 5
-
Point 21
-
Point 13
-
Point 25
-
Point 9
-
Point 18
-
Point 6
-
Point 22
-
Point 14
-
Point 26
-
Point 10
Results
-
Average of measurements (μSv/h)
-
Average below 20 μSv/h
-
All measurements lower than 100 μSv/h