Information

  • Document No.

  • Audit Title

  • Client / Site

  • Conducted on

  • Prepared by

  • Location
  • Personnel

1. Laboratory Security

Security

  • Unlocked Lab (Item of Concern): Lock laboratory doors when no one is present to secure hazardous materials and equipment. TAMU Laboratory Safety Manual, Chapter 2 Section 1.1(g), Section 1.2(a); BMBL V 2007, Section IV, Page 30 & 33

  • Add media

  • Unlocked cabinets, etc. (Item of Concern): Cabinets, refrigerators, and freezers that contain hazardous materials and are located in areas outside the laboratory must be locked. TAMU Laboratory Safety Manual, Chapter 2 Section 1.2 & Chapter 3 section 7.1(h)

  • Add media

  • Inspector not identified (Item of Concern): Visitors to lab should be approached and identified. TAMU Lab Safety Manual Chapter 2 section 1.2(g); BMBL V 2007, Section IV

  • Add media

2. Electrical Safety

Misc. Electrical.

  • No GFCI near wet areas (Deficiency): Submit request through AggieWorks to have GFCI (ground fault circuit interrupter) installed on electrical outlets located within 6 feet of wet areas. NFPA 70.210.8(B)(5)

  • Add media

  • Damaged electrical outlet (Item of Concern): Repair is needed on electrical outlet. Submit request for repair through AggieWorks. IFC 605.5.3; TAMU Safety Manual-Fire & Life Safety chapter VIII, section 11.1.5

  • Add media

  • Blocked electrical panel (Item of Concern): Remove items that block access to the electrical panel. TAMU Safety Manual - Fire & Life Safety (chapter VIII, section 11.2)

  • Add media

Extension Cords/Power Strips/Plug Adapters.

  • Permanent Use of Extension Cords (Item of Concern): Extension cords are for temporary use only (no more than 8 hours per day). Replace with fuse-protected power strip or permanent wiring. To have permanent wiring installed, submit request through AggieWorks. IFC 605.5; TAMU Safety Manual-Fire & Life Safety Chapter VIII, Section 11.1

  • Add media

  • Extension cords above ceiling tiles (Item of Concern): Remove extension cords that are placed above ceiling tiles. IFC 605.5; TAMU Safety Manual-Fire & Life Safety Chapter VIII, Section 11.1.7

  • Add media

  • Extension cords/power strips connected in series (Item of Concern): Extension cords and power strips should be plugged directly into wall receptacle, not “daisy-chained” or connected in a series. TAMU Safety Manual-Fire & Life Safety Chapter VIII, Section 11.1.3

  • Add media

  • 3-way plugs (Item of Concern): Discontinue use of 3-way plug adapters. Replace with fuse-protected power strip(s). IFC 605.4

  • Add media

  • Power strips in fume hood (Item of Concern): Remove electrical strips from fume hood. Prudent Practices 2011, Page 26 Section 2.E.3; NFPA 45

  • Add media

  • Damaged electrical cord/plug (Item of Concern): Have electrical cord or plug repaired by a certified technician. IFC 605.5.3; TAMU Safety Manual-Fire & Life Safety Chapter VIII, Section 11.1.5

  • Add media

3. Fire/Life Safety

Clearance.

  • Blocked evacuation routes (Deficiency): Keep evacuation/exit routes clear of obstructions. NFPA 101.3.3.161

  • Add media

  • Blocked sprinkler head (Item of Concern): Remove all items stored within the 18” plane below the level of the sprinkler heads, throughout the room. NFPA 1.10.19.3

  • Add media

  • Combustibles near ceiling (Item of Concern): Remove all combustible items stored within 24” of the ceiling. NFPA 1.10.19.3

  • Add media

Bunsen Burners.

  • Damaged Bunsen burner tubing (Deficiency): Replace brittle or cracked Bunsen burner tubing. ASTM D 2513 Standard

  • Add media

  • Bunsen burner in BSC (Deficiency): Do not use a Bunsen burner in a biological safety cabinet. Personal protection is reduced when the air curtain is disrupted by flame, and unburned gas can collect in filters, increasing explosion potential. (EHS will notify the Office of Research Compliance & Biosafety of this violation.) BMBL 5th edition 2007, Appendix A+B18

  • Add media

Misc. Fire/Life Safety.

  • Lab doors open (Item of Concern): Lab doors must be kept closed to maintain fire barriers and proper ventilation in lab areas. NFPA 45.8.3.3 & 45.7.4

  • Add media

  • Missing/damaged ceiling tiles (Item of Concern): Submit request through AggieWorks to replace missing/damaged ceiling tiles. NFPA 101.8.4.2(1)(2)

  • Add media

Fire Extinguishers.

  • Blocked fire extinguisher (Deficiency): Remove obstruction blocking fire extinguisher. NFPA 10.6.1.3.1 to 10.6.1.3.3

  • Add media

  • Discharged fire extinguisher (Item of Concern): Fire extinguisher has been discharged. Complete an “Accident/Investigation Form” for the incident or near-miss, and submit the form to the EHS inspector within 24 hours. (See https://ehsd.tamu.edu/Pages/OccSafety.aspx) NFPA 10.7.2.1.2

  • Add media

  • Damaged fire extinguisher (Item of Concern): Fire extinguisher is damaged. EHS inspector will notify Fire & Life Safety personnel. NFPA 10.7.2.2.2

  • Add media

4. Safe Lab Practices

Food/Beverages.

  • Food/drink in lab (Deficiency): Food and beverages may not be consumed, stored, or prepared in lab areas. 29 CFR 1910.1030(d)(2)(ix); Prudent Practices 2011, Page 109 Section 6.C.2.3

  • Add media

  • Food/drink in lab refrigerator/freezer (Deficiency): Food/beverages and chemicals may not be stored in the same refrigerator or freezer. Rather, store food and drink in an area outside of the lab. Prudent Practices 2011, Page 97 - 98, Page 109 Section 6.C.2.3 ; TAMU Laboratory Safety Manual, Chapter 2 Section 1.1

  • Add media

  • Food utensils washed in lab sink (Deficiency): Plates, cups, and silverware for food or beverage use may not be washed in the same sink as lab utensils and should not be in lab areas at all. TAMU Laboratory Safety Manual, Chapter 2 section 1.1

  • Add media

Hand-washing Facilities.

  • Sink/water needed (Item of Concern): Provide running water for hand washing. 29 CFR 1910.1030(d)(2)(iv)

  • Add media

  • Soap needed (Item of Concern): Provide soap for hand washing. 29 CFR 1910.1030(d)(2)(iv)

  • Add media

  • Paper towels needed (Item of Concern): Provide paper towels for drying hands after washing. 29 CFR 1910.1030(d)(2)(iv)

  • Add media

Sharps Handling, Storage, and Disposal.

  • Recapped needles (Deficiency): Do not manually recap needles. Use self-sheathing syringes or dispose of used syringes in a sharps waste container without recapping. 29 CFR 1910.1030(d)(2)(vii)(viii)

  • Add media

  • Unsecured sharps (Deficiency): Secure razors, knives, needles, microtome blades, and other sharps to be reused in rigid protective cases. 29 CFR 1910.1030(d)(2)(vii)(viii)

  • Add media

Housekeeping/Misc.

  • Spilled chemicals (Deficiency): Clean up spilled chemicals on bench top, in fume hood, or on floor. TAMU Laboratory Safety Manual, Chapter 3 Section 6

  • Add media

  • Clutter in lab (Item of Concern): Dispose of clutter, including unnecessary boxes, old equipment, and trash. 29 CFR 1910.22(a)(1)

  • Add media

  • Dirty fume hood (Item of Concern): Clean dirty fume hood. Prudent Practices 2011, Page 223 Section 9.C.2.5

  • Add media

  • Dirty work surface (Item of Concern): Clean dirty or untidy workspace or lab bench. TAMU Laboratory Safety Manual, Chapter 2 Section 1.4

  • Add media

5. Physical Hazards

Cryogenic Cylinders and Dewars.

  • Damaged cryogenic cylinder (Deficiency): Ribbing, excessive ice buildup on tank, cracked gaskets, excessive rust, etc. are indications that the cryogenic cylinder is in poor condition. Contact service provider to have cylinder serviced. 29 CFR 1910.101; TAMU Laboratory Safety Manual, Chapter 2 Section 2.7

  • Add media

  • Pressure relief valve compromised (Deficiency): Do not plug or otherwise compromise the pressure relief valve on a cryogenic cylinder. This is an important safety feature of the cylinder. NFPA 55.8.2.5

  • Add media

  • Blow-out disc compromised (Deficiency): Do not plug or otherwise compromise the blow-out disc on a cryogenic cylinder. This is an important safety feature of the cylinder. NFPA 55.8.2.5

  • Add media

  • Cryogenic cylinder due for maintenance (Item of Concern): TAMU-owned cryogenic cylinder is due/past due for maintenance. Obtain maintenance from service provider and affix appropriate label of current maintenance/service. 29 CFR 1910.101

  • Add media

Compressed Gas Cylinders

  • Improperly secured gas cylinder (Item of Concern): Properly secure compressed gas cylinder(s) with a device designed for such use. Ropes and bungee cords are not permitted for securing cylinders. NFPA 45.11.1.5.1; NFPA 45 Annex F.17

  • Add media

  • Gas cylinders secured in group (Item of Concern): Secure compressed gas cylinder(s) individually, not in groups. Prudent Practices 2011, Page 168 Section 7.D.3

  • Add media

  • Gas cylinder missing safety cap (Item of Concern): Apply safety cap to cylinders when not in use. NFPA 45 Annex F.14

  • Add media

Vacuum Pumps.

  • Vacuum pump missing guard (Item of Concern): Replace the belt and pulley guard that is missing from the vacuum pump. Prudent Practices 2011, Page 153 Section 7.C.2

  • Add media

  • Vacuum pump leaking oil (Item of Concern): Repair or replace the vacuum pump, which is leaking oil. Prudent Practices 2011, Page 153 Section 7.C.2

  • Add media

  • Vacuum pump venting exhaust (Item of Concern): Repair or replace the vacuum pump, which is venting exhaust into the lab. Prudent Practices 2011, Page 153 Section 7.C.2

  • Add media

  • Vacuum pump needs secondary containment (Item of Concern): Place vacuum pumps containing oil in secondary container made of non-combustible materials in order to contain any leaks that may occur.

  • Add media

Glass Equipment or Containers.

  • Unshielded glass under pressure (Item of Concern): Shield and/or wrap tape around glass that is under vacuum pressure (Rotovap, glass Dewars, etc.). Prudent Practices 2011, Page 171 Section 7.E.1.2; TAMU Laboratory Safety Manual, Chapter 2 Section 2.5

  • Add media

  • Glass items on floor (Item of Concern): Do not store glass items directly on the floor. Place in cabinets or in secondary containment. Prudent Practices 2011, Page 77 Section 4.E.10

  • Add media

Misc. Hazards.

  • Unguarded moving parts (Deficiency): Guard exposed, moving parts of equipment, including pinch points and belts. Prudent Practices 2011, Page 163 Section 7.C.8.4.2; TAMU Laboratory Safety Manual, Chapter 2 Section 2.3

  • Add media

  • Tripping hazard (Item of Concern): Relocate, tape down, or otherwise secure electrical cords, computer cables and/or hoses that present tripping hazards. Prudent Practices 2011, Page 77 Section 4.E.10

  • Add media

6. Chemical Fume Hood

Misc. Fume Hood.

  • Items on fume hood knobs (Item of Concern): Remove items that are hanging on fume hood control knobs.

  • Add media

Obstructed Fume Hood.

  • Airflow obstructed in fume hood (Deficiency): To avoid obstructing airflow in the fume hood, place items at least 6 inches back from the face of the hood and 6 inches away from the baffles in the rear of the hood. Prudent Practices 2011, Page 223 Section 9.C.2.5, Page 159 Section 7.C.5.7, Page 110 Section 6.C.2.4.1

  • Add media

  • Equipment blocks airflow in fume hood (Deficiency): Use jacks to elevate equipment 2-4” off the floor of the fume hood to improve airflow in the hood. Prudent Practices 2011, Page 223 Section 9.C.2.5, Page 110 Section 6.C.2.4.1

  • Add media

  • Clutter in fume hood (Item of Concern): Remove clutter from fume hood. Prudent Practices 2011, Page 223 Section 9.C.2.5, Page 110 Section 6.C.2.4.1

  • Add media

  • Chemicals stored in fume hood (Item of Concern): Remove chemicals stored in fume hood. Return them to chemical storage cabinets. NFPA 45.9.2.3.7; Prudent Practices 2011, Page 223 Section 9.C.2.5

  • Add media

  • Loose paper in fume hood (Item of Concern): Remove loose paper (paper towels, Kim wipes, notebook paper) from fume hood. Paper can become lodged in the ductwork, and removing it requires significant monetary costs and labor. Prudent Practices 2011, Page 223 Section 9.C.2.5, Page 110 Section 6.C.2.4.1

  • Add media

Improper Use of Fume Hood.

  • Personnel using out-of-service fume hood (Deficiency): Do not use any fume hood that has been tagged out of service. Submit work order through AggieWorks for repair. EHS must certify the repaired hood before it can be used. NFPA 45.8.13.2; TAMU Laboratory Safety Manual, Chapter 5 Section 2

  • Add media

  • Fume hood sash left up (Item of Concern): Pull fume hood sash down completely when the hood is unattended. Pull down to lowest working level when working in the hood. NFPA 45.8.8.3

  • Add media

  • Personnel using perchloric acid (70% or greater, or heated) in a regular chemical fume hood (Deficiency): Use of perchloric acid (70% or greater, or heated) in a regular chemical fume hood can lead to formation of explosive peroxides in the duct work. The department is now responsible for having the fume hood professionally cleaned before any maintenance or repair work. Contact EHS for referrals. Notify EHS at EHSD-Fumehood@tamu.edu when the fume hood has been cleaned. NFPA 45.8.11.10

  • Add media

7. Personal Protective Equipment/Eyewash/Showers

Emergency Shower.

  • Blocked emergency shower (Deficiency): Remove obstruction(s) blocking emergency shower to ensure clear access to this safety equipment. ANSI Z358.1-2009 Section 4.5.2

  • Add media

Eyewash.

  • No eyewash (Deficiency): Provide a continuous-flow (plumbed), hands-free, ANSI Z358.1 approved eyewash station. There is not one available in this lab. ANSI Z358.1-2009 Section 5

  • Add media

  • Inadequate eyewash (squeeze-bottle) (Item of Concern): Provide a continuous-flow (plumbed), hands-free, ANSI Z358.1 approved eyewash station. A squeeze-bottle eyewash is not adequate. ANSI Z358.1-2009

  • Add media

  • Inadequate eyewash (single-head drench hose) (Item of Concern): Provide a continuous-flow (plumbed), hands-free, ANSI Z358.1 approved eyewash station. A single-head drench hose is not adequate. ANSI Z358.1-2009

  • Add media

  • Eyewash unit not flushed (Item of Concern): Lab personnel must flush eyewash units once weekly for 3-5 minutes. Keep a log to record weekly activation of eyewash units (EHS will provide an Eyewash Activation Log upon request). ANSI Z358.1-2009

  • Add media

  • Non-functional eyewash (Item of Concern): Eyewash unit is in need of repair. Submit request through AggieWorks to have eyewash unit repaired. ANSI Z358.1-2009 Page 12 Section 5.5

  • Add media

Personal Protective Equipment (PPE).

  • Inadequate PPE (general) (Deficiency): Ensure lab personnel wear appropriate PPE (eye protection, lab coats, gloves, etc.) and clothing (no shorts, bare midriffs, unsecured long hair, dangling jewelry or open-toed shoes) when working in the lab. Post PPE requirements on the door to the lab based on hazards present within the lab. 29 CFR 1910.132; TAMU Laboratory Safety Manual, Chapter 5 Section 3, Prudent Practices 2011, Page 175

  • Add media

  • Safety eyewear inadequate or not worn (Deficiency): Use appropriate safety eyewear for the work being performed, such as safety glasses for impact hazards or chemical splash goggles for handling chemicals. Post PPE requirements for appropriate eyewear on the door to the lab. 29 CFR 1910.133; TAMU Laboratory Safety Manual, Chapter 5 Section 3; Prudent Practices 2011, Page 175 Section 7.F.1.3

  • Add media

  • Lab coats not worn (Deficiency): Ensure lab personnel wear lab coats or other protective clothing when working in areas with chemical or biological hazards. Post PPE requirements on the door to the lab. Prudent Practices 2011, Page 175 Section 7.F.1.1; 29 CFR 1910.132; TAMU Laboratory Safety Manual, Chapter 5 Section 3

  • Add media

  • Improper foot-wear (Deficiency): Ensure lab personnel wear closed-toe shoes in the lab. Crocs, sandals, flip flops, and other shoes that do not cover the feet are not appropriate. Post appropriate foot ware requirements on the door to the lab. Prudent Practices 2011, Page 175 Section 7.F.1.2 & 176; 29 CFR 1910.136; TAMU Laboratory Safety Manual, Chapter 5 Section 3

  • Add media

8. Chemical Safety

Flammable Chemical Storage

  • Excessive flammables outside of flammable cabinet (Deficiency): Store flammable chemicals in excess of 10 gallons in a flammable storage cabinet. NFPA 45.10; NFPA 30.9.5

  • Add media

  • Flammables stored near ignition source (Deficiency): Store flammable chemicals away from ignition and heat sources and out of direct sunlight. NFPA 45.12.2.3.3

  • Add media

  • Flammables in household-style refrigerator/freezer (Deficiency): Flammable chemicals needing refrigeration must be kept in a Lab-safe refrigerator/freezer (that is, safe for flammable storage). Flammable chemicals may not be kept in a household style or commercial refrigerator/freezer or in a walk-in cooler. NFPA 45.12.2.2.2.1

  • Add media

  • Improper storage of flammables (Item of Concern): Store flammable chemicals in flammable storage cabinet. NFPA 30.9.5

  • Add media

Chemical Dating & Retention.

  • Old non-time sensitive chemicals (Item of Concern): Noted non-time sensitive chemicals older than 5 years in the lab. Periodically review chemical inventory. Dispose of old chemicals that are no longer used. Label non-time-sensitive chemicals that are still in use with current 'in use' date.

  • Add media

  • Chloroform out of date (Deficiency): Chloroform should be used within 1 year of purchase or 6 months after opening, because it can form phosgene gas upon decomposition. Dispose of any chloroform once these time limits have passed. This is also true for chloroform that has had stabilizers added. NFPA 45.9.2.3.4

  • Add media

  • Peroxide forming chemicals out of date (Deficiency): Explosive peroxide forming chemicals and oxidants such as ethyl ether, tetrahydrofuran (THF), perchloric acid, cyclohexene, butadiene, isopropyl ether and dioxanes must be used within 1 year of purchase or 6 months after opening and must be disposed of before the expiration date. Tag these chemicals for disposal immediately if past these time constraints. NFPA 45.9.2.3

  • Add media

  • Undated chemicals (Item of Concern): Date chemicals when received and again when opened. If date of acquisition or opening is unknown, back-date to the oldest known date for a reference point, e.g. "REC'D PRE-2010" or “OPENED PRE-2010” NFPA 45.9.2.3.4; NIOSH Publication 2007-107

  • Add media

Faulty Containers & Unstable Chemicals.

  • Unstable chemicals (Deficiency): Do not attempt to open or move peroxide-forming and oxidizing chemicals (such as ethyl ether, THF, perchloric acid, cyclohexene, butadiene, isopropyl ether and dioxanes) with crust on lid, stratification of liquid, discoloration, or crystallization. (EHS will follow-up with PI).

  • Add media

  • Damaged chemical container (Item of Concern): Repackage or dispose of chemicals in deteriorated containers. Prudent Practices 2011, Page 113 Section 6.C.3

  • Add media

  • Missing or damaged chemical cap (Item of Concern): Replace cracked or missing cap on chemical container. Prudent Practices 2011, Page 113 Section 6.C.3

  • Add media

Chemical Segregation.

  • Unsegregated chemicals (Deficiency): Segregate chemicals by hazard class (flammable, corrosive, oxidizer, reactive, toxic). Recommend using a color-coded labeling system on chemical containers to quickly identify the hazard class. For more information, see the EHS Laboratory Safety webpage (https://ehsd.tamu.edu/Pages/LabSafety.aspx). NFPA 45.9.2.3.2

  • Add media

Perchloric Acid Use and Storage.

  • Improper Perchloric acid storage (Deficiency): Perchloric acid must be stored away from organic materials (including other chemicals, wooden shelves, and paper lining) and other acids. Storage in secondary containment is strongly recommended. Review MSDS for more information. NFPA 45.9.2.3; TAMU Laboratory Safety Manual, Chapter 3 Section 9.1

  • Add media

  • Using/heating Perchloric acid outside of Perchloric acid fume hood (Deficiency): A Perchloric acid fume hood is required when using 70% or greater concentration of perchloric acid or when heating any concentration of perchloric acid. NFPA 45.8.11; TAMU Laboratory Safety Manual, Chapter 5 Section 2.1

  • Add media

  • Personnel using improperly functioning Perchloric acid hood (Deficiency): A Perchloric acid fume hood must be in proper working condition, with a functioning wash down system, when using 70% or greater mixtures of perchloric acid or when heating any concentration of perchloric acid. Submit a work request through AggieWorks to have the Perchloric acid fume hood repaired. EHS must verify that the wash down system is working properly before the hood may be used. NFPA 45.8.11

  • Add media

Flammable Storage Cabinets.

  • Flammable cabinet does not vent to outside (Deficiency): If vented, flammable storage cabinet must vent directly outside the building. NFPA 30.9.5.4.2

  • Add media

  • Improper duct material on vented flammable cabinet (Deficiency): Flammable storage cabinet exhaust duct shall be constructed of non-combustible material and should be of the same material as the flammable storage cabinet. NFPA 91.4.2.1

  • Add media

  • Flammable cabinet vents from the top (Item of Concern): If vented, flammables cabinets should vent from the bottom of the cabinet, with make-up air supplied at the top of the cabinet. NFPA 30 Appendix A.9.5.4

  • Add media

General Chemical Storage.

  • Hazardous liquids above eyelevel (Item of Concern): Do not store hazardous liquid chemicals above the eye level of the shortest person working in the lab. Prudent Practices 2011, Page 95 Section 5.E.1, Page 114 Section 6.C.5

  • Add media

  • Chemicals on workbench (Item of Concern): To avoid cluttering workspaces, do not store multiple chemicals on workbenches. Return chemicals to storage cabinets when they are not actively being used. Prudent Practices 2011, Page 95 Section 5.E.1

  • Add media

  • Chemicals on edge of bench top (Item of Concern): Do not place chemical bottles on the edge of the bench top, where they can easily be knocked to the floor. Prudent Practices 2011 page 95

  • Add media

  • Heavy chemicals on upper shelves (Item of Concern): Store heavy chemical containers on the lowest shelves. Prudent Practices 2011, Page 95 Section 5.E.1

  • Add media

  • Chemicals on floor (Item of Concern): Do not store chemicals directly on the floor. Rather, store them in approved cabinets or in secondary containers large enough to hold the entire contents of the bottle in case of a spill. Prudent Practices 2011, Page 95 Section 5.E.1, Page 113 Section 6.C.3

  • Add media

9. Waste Disposal

Improper Accumulation of Waste.

  • Old samples present (Deficiency): Dispose of old samples that are no longer needed.

  • Add media

  • Full waste containers (Deficiency): Dispose of filled chemical waste containers. TAMU Laboratory Safety Manual, Chapter 4 Section 1

  • Add media

  • Used lecture bottles (Deficiency): Dispose of used lecture bottles. Prudent Practices 2011, Page 100 Section 5.E.6

  • Add media

Glass Disposal.

  • Improper glass disposal (Deficiency): Dispose of broken glass in a covered sturdy box marked "BROKEN GLASS." 30 TAC § 330.1219 (b)(4); Prudent practices 2011, Page 113 Section 6.C.3

  • Add media

  • Over-filled glass waste (Deficiency): Do not overfill glass waste container. Glass should not protrude over lip of disposal container. TAMU Laboratory Safety Manual, Chapter 4 Section 3

  • Add media

  • Unlabeled glass waste (Item of Concern): Glass disposal box should be marked "BROKEN GLASS." Prudent practices 2011, Page 114 Section 6.C.3

  • Add media

Improperly Labeled Chemical Waste.

  • Improperly labeled chemical waste (Item of Concern): Label chemical waste container with the words "HAZARDOUS WASTE" or tie a Hazardous Waste Disposal tag on the container with string. 30 TAC § 330.1207; TAMU Laboratory Safety Manual, Chapter 4 Section 1

  • Add media

  • Hazardous waste not identified (Item of Concern): List the chemical contents on the waste disposal container or on the Hazardous Waste Disposal tag tied to the container. 30 TAC § 330.1207; NFPA 45.9.3; TAMU Laboratory Safety Manual, Chapter 4 Section 1

  • Add media

  • Dated hazardous waste disposal tags (Item of Concern): DO NOT date Hazardous Waste Disposal tag. EHS will date the tag at time of pick-up. TAMU Laboratory Safety Manual, Chapter 4 Section 1.1

  • Add media

Waste Disposal.

  • Improper sharps waste container (Deficiency): Use an approved sharps container with a non-removable lid for disposal of metals sharps. If sharps are biohazardous, autoclave before disposal. If sharps are not biohazardous, remove the biohazard label from the container and label the container "NON-BIOHAZARDOUS SHARPS." When full, place the container in a black trash bag before disposing in the dumpster.

  • Add media

  • Uncapped waste (Item of Concern): Tightly cap waste container that was left uncapped. Recommend using screw-on funnel with cap for waste containers that are accessed frequently. 40CFR 265(i)

  • Add media

10. Warning Signs/Labels

Chemical Container Labels.

  • Damaged/missing chemical label (Deficiency): Replace missing or deteriorating labels on primary (original) chemical containers. 25 TAC §295.6

  • Add media

  • Improperly labeled secondary container (Item of Concern): Ensure that all containers (squeeze bottles, flasks, dilutions, etc.) are labeled with the complete name of the contents, even those that contain water or are a part of equipment or experiments. The label should also include words, pictures, and/or symbols to notify users of the hazards of the chemicals. Replace fading/deteriorated labels. TAMU Laboratory Safety Manual, Chapter 3 Section 2; 29 CFR 1910.1200(f)(9)

  • Add media

  • Improperly defaced bottles (Item of Concern): Deface labels on all empty bottles prior to reuse, and label the bottles with the identity of the new contents to prevent confusion. TAMU Laboratory Safety Manual, Chapter 4 Section 1; 25 TAC §295.6

  • Add media

Equipment Signage.

  • Unlabeled microwave (Item of Concern): Label microwave "Lab Use Only." TAMU Laboratory Safety Manual, Chapter 2 section 1.1

  • Add media

  • Unlabeled freezer (Item of Concern): Label freezer "Not Safe for Flammable Storage." Prudent Practices 2011, Page 98 Section 5.E.4; TAMU Laboratory Safety Manual, Chapter 2 section 1.1

  • Add media

  • Unlabeled refrigerator (Item of Concern): Label refrigerator "Not Safe for Flammable Storage." Prudent Practices 2011, Page 98 Section 5.E.4; TAMU Laboratory Safety Manual, Chapter 2 section 1.1

  • Add media

  • Unlabeled ice maker (Item of Concern): Label ice maker "Not for Human Consumption." TAMU Laboratory Safety Manual, Chapter 2 section 1.1

  • Add media

  • Unlabeled hot surface (Item of Concern): Place warning sign on equipment with hot surfaces. TAMU Laboratory Safety Manual, Chapter 2 Section 2.6

  • Add media

Door Signs.

  • Lack of warning signs (Item of Concern): Post specific warning signs for the hazards present (Chemicals, Biohazards, Radioactive Materials, LASER, etc.) on the lab door. (EHS inspector will notify the appropriate group with oversight.) TAMU Laboratory Safety Manual, Chapter 3 Section 3, Chapter 5 Section 1.1; 29 CFR 1910.145, 21 CFR 1040.10(g)

  • Add media

  • Improper warning signs (Item of Concern): Remove from the lab door any warning signs that are not appropriate for the hazards in the lab. (EHS will notify the appropriate group with oversight.) TAMU Laboratory Safety Manual, Chapter 3 Section 3, Chapter 5 Section 1.1

  • Add media

  • No ECI outside lab door (Item of Concern): Post emergency contacts and off-hour phone numbers on all outer lab doors. TAMU Lab Safety Manual, Chapter 5 Section 1.1

  • Add media

  • Out of date ECI (Item of Concern): Update Emergency Contact Information on lab door. TAMU Laboratory Safety Manual, Chapter 1 Section 3.2

  • Add media

11. Laser Safety

Laser Safety.

  • No Laser eyewear available (Item of Concern): EHS inspector will notify the Laser Safety Group. 25 TAC §289.301

  • Add media

  • Lab personnel have not had Laser training (Item of Concern): EHS inspector will notify the Laser Safety Group. TAMU Rule 24.01.01.M5 section 1(1.3)(1.4)

  • Add media

  • Class IIIb and Class IV Lasers not registered with EHS (Item of Concern): EHS inspector will notify the Laser Safety Group. 25 TAC §289.301

  • Add media

12. No Unsafe Conditions Found

  • No Violations Found.

13. Other Violations

  • Other: See Comments.

14. Administrative

Training.

  • Lab personnel lack general HazCom training (Deficiency): Laboratory employees must be provided general Hazard Communication and/or Introduction to Laboratory Safety Training prior to beginning work in a TAMU laboratory. This training must be documented and kept on file. THCA Section 502.009; TAMU Hazard Communication Program, Page 5, Chemical Safety Information And Training - HSC 502.009 and 502.017(b) and Page 6; 25 TAC §295.7(b)

  • Add media

  • Lab personnel lack work area specific training (Deficiency): Laboratory employees must be provided Work Area Specific Training prior to beginning work in a TAMU laboratory and whenever new hazards, new equipment, and/or new processes are introduced. This training must be documented and kept on file. THCA Section 502.009; TAMU Hazard Communication Program, Page 5, Chemical Safety Information And Training - HSC 502.009 and 502.017(b) and Page 6; 25 TAC §295.7(b)

  • Add media

  • Lab personnel lack hazard specific training (Deficiency): Laboratory employees must be provided Work Area Hazard Specific Training for extremely toxic chemicals (such as Hydrogen Fluoride, Phosgene), explosives, pyrophorics, etc., prior to beginning work in a TAMU laboratory and whenever new hazards, new equipment, and/or new processes are introduced. This training must be documented and kept on file.

  • Add media

Documentation.

  • No MSDS available (Deficiency): A current manufacturer-specific safety data sheet (MSDS or SDS) must be maintained and made available to lab employees for each chemical used or stored in the lab. THCA Section 502.006; TAMU Hazard Communication Program, Page 6 - Material Safety Data Sheets - HSC 502.006; 25 TAC §295.5

  • Add media

  • Lab not decommissioned by EHS (Deficiency): Laboratories must be decommissioned by Environmental Health & Safety whenever a lab will be permanently closed or will be vacated by the current faculty member/principal investigator (PI). The decommissioning process, including a close-out inspection, must be completed prior to the new PI moving into the lab. TAMU Rules: (SAP) 24.01.01.M4.04

  • Add media

  • Lab not registered with EHS (Item of Concern): The PI has not registered this lab with EHS. Complete the “PI Laboratory Registration Form” (found on https://ehsd.tamu.edu/Pages/LabSafety.aspx) and submit to EHS within 5 business days.

  • Add media

The templates available in our Public Library have been created by our customers and employees to help get you started using SafetyCulture's solutions. The templates are intended to be used as hypothetical examples only and should not be used as a substitute for professional advice. You should seek your own professional advice to determine if the use of a template is permissible in your workplace or jurisdiction. You should independently determine whether the template is suitable for your circumstances.