Title Page
-
Conducted on
-
Prepared by
-
Location
Water Activity Meters
-
Use the 0.76 standard first then the 0.50
-
Specifications: 0.76 +/- 0.02, 0.50 +/- 0.02
-
Water Activity is KDR's CCP #2. All meters must be calibrated at least once every day.
Meter Information
-
Serial Number
-
0.76 standard results
-
Verification failed. Clean meter and make adjustments then record retest results
-
Remove meter from service for cleaning and/or repair.
-
0.50 standard results
-
Verification failed. Clean meter and make adjustments then record retest results.
-
Verification failed. Remove meter from service for cleaning and/or repair.
pH Meter
-
Using the 4.00 & 7.00 standards, record the slope percentage results. Slope must be between 90 - 105%
-
Calibration failed. Clean probe and replace standards. Record retest results.
-
Calibration failed. Remove from service for repair.
Scale
-
Using the 200g weight, record digital read out results.
-
Calibrate scale and record retest results
-
Calibration failed. Remove from service for repair.
Pipette-100ul
-
Pipette one pump (100ul) into tared cup on a calibrated scale scale. Results must be 0.1g
-
Verification failed. Make adjustments to pipette and record results.
-
Calibration has failed. Remove from service for repair.
Pipette-500ul
-
Pipette one pump (500ul) into a tared cup on a calibrated scale. Results must be 0.5g
-
Verification failed. Make adjustments to pipette and record results.
-
Calibration has failed. Remove from service for repair.