Title Page
-
Site conducted
-
Conducted on
-
Prepared by
-
Location
Compliance with
-
Precleanse the flexible endoscope at point of care.
-
Transports the contaminated endoscope and accessories to the endoscopy processing room as soon as possible after use.
-
Ensures the container is labeled with a biohazard label.
-
Ensures the container was cleaned with Sani-wipe and the IFU followed (wet time).
-
The area under the sink are free of dust and debris
-
The staff member is wearing the proper PPE appropriately
Manual Cleaning
-
Performs leak testing before manual cleaning.
-
GI endoscope reprocessing staff performed daily flow validation testing on endoscope flushing device
-
Completely submerges the endoscope and accessories.
-
Cleans all accessible channels and the end of the endoscope with a cleaning brush of the length, width, and material recommended by the manufacturer.
-
Uses a clean disposable brush for each endoscope cleaning.
-
If the endoscope has an elevator, raises and lowers the elevator throughout the manual cleaning process.
-
Flushes the channels of the endoscope with the cleaning solution.
-
Checks the expiration date of the high level disinfectant or liquid chemical sterilant before each use.
-
Uses a test strip or other FDA-cleared testing device specific to the disinfectant or liquid chemical sterilant before each use.
-
Can state the time associated with each strip (submerge for 1 second, read in 75 seconds)
-
SPD HLD GI flexible endoscopes, test strips used to verify effectiveness of disinfectant used in automated endoscope reprocessing devices are dated with a revised expiration date upon opening, contrary to strip IFU.
-
Positions endoscopes and accessories within the mechanical processor to ensure contact of the processing solutions.
-
GI staff use the enzymatic cleaner per the IFU and the temperature of the water is 68-104 F.
-
Wears clean gloves when transporting the endoscope to and from the storage cabinet.
-
Stores flexible endoscopes horizontally or hangs them vertically so they do not coil or touch the floor of the cabinet
Records
-
Processing records include: date and time
-
identity of endoscope and endoscope accessories
-
method and verification of cleaning and results of cleaning verification testing
-
number or identifier of the mechanical processor or sterilizer and results of process efficacy testing
-
identity of the persons performing the processing
-
lot number of the processing solutions
-
disposition of defective items or equipment
-
maintenance of water systems, endoscopes and endoscope accessories, and processing equipment.
Endoscope Pre-Op and PACU
-
In GI nurse station glucometer kits contained reagents and are labeled with 3 months use by date.
-
All other areas are observed compliant.