Information
-
Document No.
-
Client / Site
-
Conducted on
-
Prepared by
-
Which gas(es) does the customer have onsite?
- Carbon Dioxide
- Nitrogen
- CO2/N2 mixture
-
Is the correct monitor being used for the correct gas
-
Is the monitor installed at the correct height?
-
Is the monitor installed close to the source of the gas supply?
-
Is the monitor installed where the gas is being used?
-
Is a Cellaguard Repeater Unit mounted at every point of entry to the room where the monitor is installed?
-
Has a Splashguard been installed on all monitors (and repeaters where applicable)?
-
Has a Relay Connection Box been installed?
-
If there is a Relay Connection Box is it installed adjacent to the monitor?
-
Has an isolator been hard wired to the monitor and fitted with a security seal?
-
Have all of the required signs been erected at the points of entry to the room where the monitor is installed?
- Restricted Entry Sign
- Green Diamond
- Caution Sign
- CO2/O2 Warning Sign
-
Has all the commissioning paperwork been completed including a basic instrument check and verification of alarm settings?
-
Has the customer handover procedure been followed and recorded?
-
Has fixed installation pricing been used as detailed in SRM?