General
-
Competency checklists with documentation training for each person eligible to perform High Level Disinfection is on file and available in the room?
-
Appropriate High Level Disinfection policies are readily available in the room and signed by Medical Director.
-
Are manufacturer maintenance and instrument manuals are maintained and are available in the room?
-
Are instruments pre-cleaned per manufacturers guide<br>lines and allowed to fully dry prior to disinfecting?
-
Appropriate PPE and Spill Kits are available?
-
Daily High Level Disinfection Probe<br>Instrument Log is filled out <br>and up to date.
-
High Level Disinfection Solution Log <br>is filled out and up to date.
-
Is eye wash available and checked<br> weekly with documentation?
-
Are there any expired solutions?
-
Is a timer available and utilized?
-
Is a thermometer available and temperature checked prior to each use?
-
All Instrument processing containers<br>have a lid that is secure and labeled <br>with a BIOHAZARD and DIAMOND Label.
-
Only manufacturer recommended brushes are utilized. Brushes are disinfected following manufacturer guidelines. They should not be lying on the sink.
-
Are chemical and or biohazard materials disposed of properly?
-
Equipment has PM stickers current within<br>the last 12 months and appropriately <br>calibrated?
-
No Food, Drinks, or Make up are stored or applied in or around the designated HLD area.
-
Room Temperature is maintained between 59-86F for proper storage CidexOPA, Revital-Ox Solutions and/or Trophon Cartridges.
GUS
-
Only approved solutions are being utilized in the GUS System. <br>(Cidex OPA or Revital-Ox)
-
Enzol Instructions for use <br>with GUS are posted.
Cidex OPA
-
HLD with Cidex OPA <br>Understanding the Process<br>Educational Slides are available
-
Are quality Control test performed with each new bottle?
-
Concentration test of MEC performed prior to each use with Cidex OPA test strips? (pass/Fail)
-
Cidex- OPA Test Strips are discarded 90 days from opening date?
-
Cidex OPA solutions is labeled with date opened and 75 day expiration date.
-
Soaking container is labeled with date solution was poured and 14 day expiration date.
-
Soak time of 12 minutes is observed for each instrument.
-
Are three (3) 1 minute separate rinses performed with sterile or potable (drinking) water?
-
Quality Control Instructions <br>are posted.
-
Cidex OPA High Level Disinfection <br> Instructions are posted.
Revital-Ox
-
HLD with Revital-ox Process <br>Educational Slides are available<br>for reference in the room.
-
Are quality Control test performed with each new bottle?
-
Concentration test of MEC performed prior to each use with Revital-Ox R60 test strips? (pass/Fail)
-
Revital-Ox Resert R60 solution test strips are discarded 90 days from opening date?
-
Revital-Ox solution is labeled with the date opened and the 90 day expiration date.
-
Soaking container is labeled with date solution was poured and 21 day expiration date.
-
Soak time of 8 minutes is observed for each instrument.
-
Is a single rinse with sterile or potable (drinking) water performed with each instrument?
-
Quality Control Instructions are posted.
-
Revital- Ox High Level Disinfection <br>Instructions are posted.
Trophon
-
Trophon HLD Process <br>Educational Slides are available <br>for reference in the room.
-
Trophon EPR left connected to power and switched on at all times?
-
Chemical Indicator available and unexpired? Stored away from direct UV light?
-
Cartridges changed every 30 days
-
Single use cleaning clothes available for cleaning before disinfecting?
-
Quality Control is interpreted at end of each cycle and documented.
-
Quality Control Instructions are posted.