1. Laboratory Classification- Requirement to complete an additional checklist.
-
Does this laboratory have an OGTR,PC or QC classification? <br>If so an additional inspection checklist must be completed to ensure compliance with relevant legislation.<br>other checklists available for specific laboratories:<br>OGTR PC2 Laboratory Checklist<br>OGTR PC2 Constant temperature room checklist<br>OGTR PC2 Arthropod Facility checklist<br>OGTR PC2 Plant Facility Checklist<br>OGTR PC2 Animal Facility checklist<br>OGTR PC2 Aquatic Organism checklist<br><br><br>
1.1 Background information for the auditor
-
Are there any outstanding OHS issues arising from previous audits of this laboratory that have not been addressed<br>If so make a note at the appropriate audit question that this is an ongoing .You may attach additional information to the report.
2.1 Emergency Procedures
-
Emergency procedures and floor plans are clearly displayed.
-
Occupants are aware of emergency procedures<br>Note: Are there specific emergency procedures for this laboratory? Are occupants aware of the emergency procedures? (exposure to HF, low oxygen alarm, quenching of liquid nitrogen into laboratory, chemical spill)
-
Emergency contacts are clearly displayed and current.
-
Emergency exits are clear.
-
Emergency lighting is provided in:<br> - areas a where there is no natural lighting<br> - escape routes, including stairwells<br> - public areas<br>
2.2 Egress and evacuation
-
Exit doors are marked and clearly visible
-
Exit doors can be opened from the inside ( no padlocks)
-
Exit corridors and doors are clear of obstructions, including stored materials or equipment.
-
Exit ladders and catwalks are clear of obstructions
-
Exit lights are provided, including directional indicators where required
2.3 First aid
-
A first Aid kit is available
-
Lists of trained First Aiders and contact numbers are displayed and current
-
Kits are inspected on a regular basis
-
Cabinets and contents are clean and properly stocked. <br>List date of last check.
2.4 Fire extinguishers
-
Fire extinguishers are readily accessible in easy to see locations.
-
Fire extinguishers have been inspected and tagged within the last six months. <br>List the date of last inspection as noted on the tag.
-
Overhead sprinkler/detectors are clear of obstructions and clear of obvious visible damage.
2.5 Floors and aisles
-
The floor surface is even ( no cracks or holes), openings in the floor are protected
-
The area is free of tripping hazards.<br>Floors and aisles are clear of rubbish, materials and equipment.<br>No electrical cords are lying across walkways
-
There is sufficient width of aisles for movement ( at least 1 M clearance )
-
There are no worn or broken treads on stairs.
-
Hand rails are in good repair
-
Non-skid strips on stairs are in good condition
-
Landings are clear of obstructions
-
Emergency exit stairs are adequately lit
-
There is no storage of materials or equipment in emergency exit stairs
2.6 Environment
-
Lighting levels are sufficient and not excessive for general tasks, movement throughout the workplace and for tasks that that have specific lighting requirements
-
Glare from both external and internal light is controlled
-
Light fittings including diffusers, reflectors, tubes, bulbs, are clean and in good condition
-
Noise exposures is prevented<br> where these is noisy equipment or processes a noise survey has been conducted to ascertain required controls (including substitution, isolation, & hearing protection)<br>Contact the OHS consultant to request a noise assessment if required.
-
General ventilation , temperature and humidity in the area appear sufficient<br>
2.7 Handling and storage of material
-
Mechanical aids or equipment are available for handling materials
-
Materials are stored on shelves or in storage rooms as appropriate.
-
Heavy items are not stored above shoulder height
-
Storage material is secured to prevent shifting/falling.<br><br>
-
Storage units are free of rubbish and properly secured
-
Storage units are properly constructed and secured to prevent shifting/falling
-
Storage rooms are neat and tidy
-
Adequate storage is proved for PPE
3.1 Laboratory facilities (general)
-
Access to the laboratory is restricted to appropriate personnel
-
Appropriate PPP is available and in use ( lab coat, safety glasses, fully enclosed footwear)<br>Note: Check if any other PPE is used in the lab. Is it appropriate for the tasks conducted?
-
Safe work procedures have been prepared for hazardous equipment, chemicals and processed and are readily available ( e.g posted next to equipment,in lab safety folder)
-
Adequate washing facilities are provided
-
Food and drink are not stored r consumed in the laboratory
-
Warning signs for radiation, biological or other hazards are prominently displayed and in good condition.
-
Spill kits, supplies and decontamination material are available for the hazards present.
-
Spill procedures are clearly displayed.
-
Appropriate PPE is available and in use
3.2 Waste management
-
All waste is disposed of appropriately
-
Hazardous or EPA prescribed wastes contained within bunded areas that do not drain to storm water or sewer
-
Hazardous waste is clearly labeled with waste description, biohazard symbols if required and date produced. The waste label Includes the waste generator contact details (name, department/school/unit and contact number) and if the waste has had any on-site treatment or testing
-
Wastes are segregated according to hazard, treatment requirements and to optimize recycling opportunities
-
Wastes are stored in "fit for purpose" containers that are able to withstand storage conditions and transport
-
Wastes are stored at appropriate temperatures
-
There are sufficient and appropriate waste disposal containers
-
Wastes are disposed of regularly and volumes are kept to a minimum
3.3 Electrical equipment
-
Laboratory equipment, including cords and leads, are in a good working order
-
There are sufficient general purpose outlets (GPOs) for the equipment
-
Double adapters and piggy-back plugs are not in use
-
Extension leads are only used for temporary, short-term power connections
-
Power boards with overload protection are used if sufficient GPOs are not available
-
Fittings are protected against external damage
-
Equipment and appliances are earthed or double insulated. Look for the double insulated symbol on the specification plate
-
Equipment that may be left on unattended has a manually resetting oven-temperature cut-off switch fitted
-
Electrical equipment is tested at appropriate intervals
3.4 .Special equipment
3.4.1 Refrigerators / freezers / cool rooms,
-
Seals on doors are in good condition
-
Refrigerators, freezers, cool rooms are appropriately labelled
-
Freezers and cool rooms have well maintained emergency door releases
3.4.2 Fume cupboards
-
Fume cupboards are appropriate for the type of hazard (i.e. radiation, biological)
-
The fume cupboard is used with the sash at or below the recommended height
-
Flammable liquids are only stored in fridge/freezers that are intrinsically safe
-
The leading edge of the fume cupboard is clear of equipment and other items
-
The area behind the main work area is clear
-
The fume cupboard is free of clutter and large equipment
-
The doors and windows within 1m of the cupboard are closed when the cupboard is used
-
Emergency switches are clearly identified for power and gas supply
-
The fume cupboard has been inspected and certified within the last 12 months
3.4.3 Biosafety cabinets
-
The Biosafety cabinet is clean and free of clutter ( i.e. routine cleaning performed recently)
-
The Biosafety cabinet has been inspected and certified within the past 12 months
3.5 Chemicals (general)
3.5.1 Chemical management
-
A register of chemicals is available
-
MSDS are available for chemicals
-
Documented risk assessments are available for chemical use
-
Carriers are available for large containers (>2L)
3.5.2 Labelling of chemicals
-
Unattended reactions are clearly labelled
-
Chemicals in use or storage labelled clearly and unambiguously
-
All reused containers have label removed or completely covered
-
Decanted chemicals are labelled
3.5.3 Chemical storage
3.5.3.1 Storage on open shelves
-
All containers are <5kg or 5L
-
On shelves >1.5m high all glass containers are <1kg or 1L
-
The shelving is chemically resistant to the chemicals stored
3.5.3.2 Storage in fire-proof cabinets
-
All containers in cabinets are <25 L in size
-
The self-closing mechanisms of the door are operating correctly
-
Incompatible chemicals stored in the same cabinet have separate spill containers
3.5.3.2 Storage in ventilated cupboards
-
All containers in cupboard are <2 kg or 25L in size
-
Incompatible chemicals stored in the same cabinet have separate spill containment
-
Ventilation is sufficient to prevent build up of offensive odors
3.6 Chemicals (specific)
3.6.1 Compressed & fuel gases
-
All the cylinders inside the room are connected for use ( no laboratory storage)
-
Gas cylinders are each securely chained with individual chains secured around the body of the cylinder and the chain ends attached to 2 separate securing points
-
Cylinders, pipes & valves are protected from mechanical damage
-
Strength of hoses and clamps attached to gas cylinder regulators are sufficient for maximum regulator delivery pressure
-
Flashback arresters are used I fuel gas supply lines
-
Non-return (check) valves are fitted to all processes that exceed cylinder pressure
-
There is a cylinder trolley available for carrying cylinders
3.6.2 Cryogenic liquids
-
Cryogenic liquids are located In well ventilated areas
-
Containers used for cryogenic liquids are designed for that purpose
3.7 Biological hazards
-
Documented risk assessments for use of biological hazards are available
-
Laboratory doors are closed when work is in progress
-
Hand washing facilities are available
-
Biological material are clearly labelled, dated and appropriately stored
-
Appropriate disinfectants are available
-
Appropriate waste handling procedures are in place (autoclaving, disinfecting etc.)
-
Biohazardous wastes and sharps bins are provided, where necessary
-
Separate facilities are provided for animal storage, cleaning, examination etc.
-
No items are stored on the floor
3.8 Ionising radiation hazards
3.8.1 unsealed sources
-
Documented risk assessments and safe work procedures for use of ionizing radiation are available
-
Work surfaces are impervious (wooden benches and tiled surfaces are unsuitable)
-
Hand washing facilities are available
-
Appropriate shielding material is available
-
Secondary containment is available and utilized (i.e. spill trays)
-
All sources of ionizing radiation are labelled and stored in secure locations
-
All storage locations of ionizing radiation are labelled
-
Appropriate monitoring equipment is available
-
Decontamination/emergency procedures are clearly displayed
3.8.2 Sealed Sources
-
Documentmented risk assessments and safe work procedures for use of ionizing radiation are available
-
Appropriate shielding material is available
-
Hand washing facilities are available
-
All sources of ionizing radiation are labelled and stored in secure locations
-
All sources of ionizing radiation are labelled and stored in secure locations
-
All storage locations of ionizing radiation are labelled
-
Appropriate monitoring equipment is available
-
Sealed radiation sources are maintained in good condition
3.8.3 x-ray apparatus or sealed apparatus
-
Documentmented risk assessments and safe work procedures for use of ionizing radiation are available
-
Appropriate shielding material is available
-
Appropriate monitoring equipment is available
-
Appropriate fail-safe devices to prevent accidental exposure are in place
-
Appropriate Labelling (including registration labels) are in place
3.9 Ultraviolet radiation
-
Appropriate shielding &/or interlocks are in place (eg interlocks on germicidal lamps; and protective shields on transilluminators)
-
Appropriate PPE is available (i.e. additional skin and eye protection may be required)
3. 10 Lasers
-
Access to Class 3 and Class 4 lasers is controlled
-
Beam stops are installed and access to the beam is restricted
-
Appropriate fail-safe devices to prevent accidental exposure are in place
-
Documented risk assessments and safe work procedures for use of lasers are available
-
Warning signs are clearly displayed
-
Wavelength-specific eye protection is available for laser operators
4.0 Environmental issues
-
Chemical containers are cleaned prior to Recycling or disposal<br>- triple rinsed with labels and lids removed before placement in the bottle and can recycling bins<br>- containers with oily or persistent residues are washed with detergent and labels removed before placing in the bottle and can recycling bins
-
Contaminated glassware that is not cleaned as above is treated as prescribed industrial waste
-
There is no evidence of incorrect disposal down sinks, staining, odorous sinks etc.
-
There are no odours, dust or smoke in area
-
All service areas, where chemicals are used and silenced or where wastes are stored are suitably bunded/protected from leakage into storm water or sewer drains
-
All electrical equipment, including computers. Are switched off when appropriate
-
Lights out stickers are placed on light switches