Title Page

  • Hospital Visited

  • Unit/Ward

  • Conducted on

  • Auditor

  • Location

Crash Cart Files

  • Crash cart folder available, complete and updated with files arranged in accordance to the KFMC Crash Cart policy,

  • Specimen signature log of all staff in the unit allocated to check the crash cart is complete and up-to-date

  • Completed checklist and test strips are kept in the folder for a period of one week- old checklist (3 months) are kept in the Unit Manager's office

  • Crash cart or equipment checked once per shift using the appropriate trolley checklist- documentation of crash cart checking complete and kept in crash cart folder (once per day for areas that do not operate 24 hours/day)- per shift for outside contents and per month of all crash cart items

  • For items in the checklist marked as (x), the staff's ID, signature and the date is documented together with the corrective action to be taken on the crash cart report

  • Crash cart deficiencies are entered in the Crash cart deficiency report- all deficiencies not resolved within 24 hours should be reported through datix

General Crash Cart Inspection

  • Crash cart is in place and accessible

  • Crash cart is locked with the number of the lock recorded on the checklist per checking-integrity of lock checked

  • Drawers are properly labeled (airway, drugs, breathing, circulation)

  • Crash cart is clean and functional- PPM up-to-date

Crash Cart Equipment Checking

  • Defibrillator available, clean and disinfected- functionality checked in accordance with the operator's manual- PPM up-to-date

  • Defibrillator plugged into an emergency power source

  • Defibrillator cables are connected to the paddles

  • Portable suction machine available, functionality checked-plugged into emergency power source- PPM up-to-date

  • Oxygen cylinders checked-have at least 1000 psi

  • Resuscitation bag connected to oxygen cylinder's flow meter using clean oxygen tubing-resuscitation bag connected to a mask

Crash Cart Content Checking

  • All crash cart items are checked for the integrity of their packaging and expiration date

  • No expiring items within the next month

  • All crash cart items are placed in their correct drawer

  • Laryngoscope (functionality) is checked once per month-with spare battery- expiration date of battery checked

  • Medication tray checked by pharmacist once a month- documented in crash cart checklist

  • Crash cart locks are locked in cupboard separate to the crash cart

undefined

  • Add signature

The templates available in our Public Library have been created by our customers and employees to help get you started using SafetyCulture's solutions. The templates are intended to be used as hypothetical examples only and should not be used as a substitute for professional advice. You should seek your own professional advice to determine if the use of a template is permissible in your workplace or jurisdiction. You should independently determine whether the template is suitable for your circumstances.