28.1 - CALIBRATION (BASE)

  • Measuring and monitoring equipment critical to product safety, legality and quality must be checked for accuracy at pre-determined frequencies

  • A contingency or back-up device must be available in the event of CCP / legal measuring equipment being out of service or away for repair

28.2 - CALIBRATION (BASE)

  • All measuring and monitoring equipment must be calibrated to national standards where possible or have documented calibration records traceable to national standards

  • WGLL - temperature probes are externally calibrated to national standards annually and then verified internally throughout the year

  • WGLL - alternatively, a reference probe is externally calibrated to National Standards annually and factory temperature probes are verified against the probe at a defined frequency

28.3 - VERIFICATION (BASE)

  • Verification checks across the normal operating range must be conducted on calibrated equipment that is critical to food safety and legality, based on risk assessment

  • All portable / handheld CCP equipment is verified on a daily basis e.g temperature probes

28.4 - MASTER LIST (BASE)

  • A master list / calibration matrix of all measuring and monitoring equipment requiring calibration must be maintained

  • WGLL - the master list / calibration matrix may include;

  • - serial numbers and or ID numbers of equipment

  • - frequency of calibrations

  • - date of last calibration

  • - date next calibration due

  • - frequency of internal verification

  • - acceptable equipment tolerances

28.5 - PROCEDURES (BASE)

  • Measuring and monitoring equipment must be calibrated / verified for accuracy against written procedures detailing:

  • - frequency of calibration / verification

  • - method of calibration / verification

  • - Acceptable equipment tolerances

  • - corrective action to be taken if outside tolerance

28.6 - CALIBRATION RECORDS (BASE)

  • Calibration certificates and records of verification must be available and up to date

28.7 - EQUIPMENT (BASE)

  • Measuring and monitoring equipment must be protected from unauthorised adjustment, damage, deterioration and misuse

28.8 - EQUIPMENT (BASE)

  • Equipment that is opening outside of specified limits must be take out of service, replaced or sent for repair

  • Documented corrective action must be evident where inaccurate measuring or monitoring equipment has been used

28.8.1 - EQUIPMENT (ASPN)

  • Where equipment is out of service or away for repair a back up device is available for use

28.9 - TEMPERATURE PROBES (BASE)

  • Temperature probes must be calibrated / verified at the temperature range at which they are used

28.10 - WEIGHTS (BASE)

  • Standard weights that are used for verification must be of the same weight range as the products being produced

28.11 - EQUIPMENT (BASE)

  • Equipment e.g. Flow meters, counting devices, timer devices and ovens etc. must be calibrated at a frequency at least as recommended by the equipment manufacturer

28.12 - TRAINING (BASE)

  • All calibration / verification must be carried out by trained personnel only

The templates available in our Public Library have been created by our customers and employees to help get you started using SafetyCulture's solutions. The templates are intended to be used as hypothetical examples only and should not be used as a substitute for professional advice. You should seek your own professional advice to determine if the use of a template is permissible in your workplace or jurisdiction. You should independently determine whether the template is suitable for your circumstances.