28.1 - CALIBRATION (BASE)
-
Measuring and monitoring equipment critical to product safety, legality and quality must be checked for accuracy at pre-determined frequencies
-
A contingency or back-up device must be available in the event of CCP / legal measuring equipment being out of service or away for repair
28.2 - CALIBRATION (BASE)
-
All measuring and monitoring equipment must be calibrated to national standards where possible or have documented calibration records traceable to national standards
-
WGLL - temperature probes are externally calibrated to national standards annually and then verified internally throughout the year
-
WGLL - alternatively, a reference probe is externally calibrated to National Standards annually and factory temperature probes are verified against the probe at a defined frequency
28.3 - VERIFICATION (BASE)
-
Verification checks across the normal operating range must be conducted on calibrated equipment that is critical to food safety and legality, based on risk assessment
-
All portable / handheld CCP equipment is verified on a daily basis e.g temperature probes
28.4 - MASTER LIST (BASE)
-
A master list / calibration matrix of all measuring and monitoring equipment requiring calibration must be maintained
-
WGLL - the master list / calibration matrix may include;
-
- serial numbers and or ID numbers of equipment
-
- frequency of calibrations
-
- date of last calibration
-
- date next calibration due
-
- frequency of internal verification
-
- acceptable equipment tolerances
28.5 - PROCEDURES (BASE)
-
Measuring and monitoring equipment must be calibrated / verified for accuracy against written procedures detailing:
-
- frequency of calibration / verification
-
- method of calibration / verification
-
- Acceptable equipment tolerances
-
- corrective action to be taken if outside tolerance
28.6 - CALIBRATION RECORDS (BASE)
-
Calibration certificates and records of verification must be available and up to date
28.7 - EQUIPMENT (BASE)
-
Measuring and monitoring equipment must be protected from unauthorised adjustment, damage, deterioration and misuse
28.8 - EQUIPMENT (BASE)
-
Equipment that is opening outside of specified limits must be take out of service, replaced or sent for repair
-
Documented corrective action must be evident where inaccurate measuring or monitoring equipment has been used
28.8.1 - EQUIPMENT (ASPN)
-
Where equipment is out of service or away for repair a back up device is available for use
28.9 - TEMPERATURE PROBES (BASE)
-
Temperature probes must be calibrated / verified at the temperature range at which they are used
28.10 - WEIGHTS (BASE)
-
Standard weights that are used for verification must be of the same weight range as the products being produced
28.11 - EQUIPMENT (BASE)
-
Equipment e.g. Flow meters, counting devices, timer devices and ovens etc. must be calibrated at a frequency at least as recommended by the equipment manufacturer
28.12 - TRAINING (BASE)
-
All calibration / verification must be carried out by trained personnel only