Title Page
-
Step 1 - Select an approved and calibrated test lamp, voltage indicating device or tester.
-
Step 2 - Verify that the device is functioning correctly on a known supply or proving unit
-
Step 3 - Locate and identify the circuit or equipment to be worked upon
-
Is the circuit or equipment in service? (Any equipment seen or known to be currently powered on the circuit)
-
Step 4 - Identify the means of isolation
-
Step 5 - Ensure isolation of the circuit or equipment by 1- Switching Off, 2- Withdrawing Fuses, 3- Locking off isolation switches or MCBs.
-
Step 6 - Verify that the circuit or equipment to be worked upon is dead using a voltage indicating device or tester, testing between 1- All Phases to Earth, 2- All Phases to Neutral, 3- Neutral to Earth
-
Discover why with care, go back to Step 4 "Identify means of isolation"
-
Fit warning labels
-
Re-check that the voltage indicating device or test equipment is functioning correctly on a known supply
-
Begin the work
-
Replace or repair the test equipment, use a proven unit for the works
-
Establish where and why the circuit is dis-energised
-
Replace or repair equipment, do not proceed with the equipment until the test equipment is proven ok.