Information
-
Document No.
-
Audit Title
-
Client / Site
-
Conducted on
-
Prepared by
-
Location
-
Personnel
Anaesthetic machine
-
Medical gas supply checked to theatre
-
Pipeline gas attached securely
-
Pipeline gas pressure 3-4 ATM
-
Oxygen cylinder pressure > 100,000 kPa
-
Gas supply alarm checked (oxygen alarm)
-
Vapourisers seated, interlock checked and full
-
All rotameters workings and bobbins spinning
-
Antihypoxic device functional (nitrous/oxygen ratio)
-
Back bar check (Negative pressure check or bouncing bobbin)
-
Circuit pressurises to 40cm H2O with no fall in 5 seconds
-
APL pop-off valve works
-
APL valve opens rapidly and pressure remains low with high flow
-
Inspiratory and expiratory valves working
-
Oxygen flush works
-
Ventilator works with test lung and minimal gas flow
-
Soda lime canister colour normal
-
Scavenging working (-75mmHg)
Monitoring, airway and other equipment
-
Monitoring turned on
-
Gas monitoring functional
-
Suction working
-
Two laryngoscopes
-
ETT several sizes, LMA several sizes
-
Oral and nasopharyngeal airways
-
Lubrication
-
Face masks, varying sizes
-
Cricothyroidotomy kit and cricothyroid cannula equipment present
-
Intubating bougie and stylet present
-
Self inflating bag and mask present
-
Scissors, clamp and Magill forceps present
-
Tapes, ties
-
Monitoring cables present
Drugs
-
Salbutamol present
-
Suxamethonium present
-
Metaraminol, ephedrine, adrenaline, atropine present